Prior to placing a device on the market, manufacturers shall undertake an assessment of the conformity of that device, in accordance with the applicable conformity assessment procedures set out in Annexes IX to XI.
Manufacturers of class IIb devices, other than custom-made or investigational devices, shall be subject to a conformity assessment as specified in Chapters I and III of Annex IX, and including an assessment of the technical documentation as specified in Section 4 of that Annex of at least one representative device per generic device group.
However, for class IIb implantable devices, except sutures, staples, dental fillings, dental braces, tooth crowns, screws, wedges, plates, wires, pins, clips and connectors, the assessment of the technical documentation as specified in Section 4 of Annex IX shall apply for every device.
Where justified in view of well-established technologies, similar to those used in the exempted devices listed in the second subparagraph of paragraph 4 of this Article, being used in other class IIb implantable devices, or where justified in order to protect the health and safety of patients, users or other persons or other aspects of public health, the Commission is empowered to adopt delegated acts in accordance with Article 115 to amend that list by adding other types of class IIb implantable devices to that list or removing devices therefrom.
6.
Manufacturers of class IIa devices, other than custom-made or investigational devices, shall be subject to a conformity assessment as specified in Chapters I and III of Annex IX, and including an assessment of the technical documentation as specified in Section 4 of that Annex of at least one representative device for each category of devices.
Alternatively, the manufacturer may choose to draw up the technical documentation set out in Annexes II and III coupled with a conformity assessment as specified in Section 10 or Section 18 of Annex XI. The assessment of the technical documentation shall apply for at least one representative device for each category of devices.
7.
Manufacturers of class I devices, other than custom-made or investigational devices, shall declare the conformity of their products by issuing the EU declaration of conformity referred to in Article 19 after drawing up the technical documentation set out in Annexes II and III. If those devices are placed on the market in sterile condition, have a measuring function or are reusable surgical instruments, the manufacturer shall apply the procedures set out in Chapters I and III of Annex IX, or in Part A of Annex XI. However, the involvement of the notified body in those procedures shall be limited:
(a)
in the case of devices placed on the market in sterile condition, to the aspects relating to establishing, securing and maintaining sterile conditions;
(b)
in the case of devices with a measuring function, to the aspects relating to the conformity of the devices with the metrological requirements;
(c)
in the case of reusable surgical instruments, to the aspects relating to the reuse of the device, in particular cleaning, disinfection, sterilization, maintenance and functional testing and the related instructions for use.
8.
Manufacturers of custom-made devices shall follow the procedure set out in Annex XIII and draw up the statement set out in Section 1 of that Annex before placing such devices on the market.
In addition to the procedures applicable pursuant to paragraph 3, 4, 6, or 7 of this Article, in the case of devices referred to in the first subparagraph of Article 1(8), the procedure specified in Section 5.2 of Annex IX or Section 6 of Annex X, as applicable, shall also apply.
10.
In addition to the procedures applicable pursuant to paragraph 3, 4, 6, or 7 of this Article, in the case of devices that are covered by this Regulation in accordance with point (f) or (g) of Article 1(6) and with the first subparagraph of Article 1(10), the procedure specified in Section 5.3 of Annex IX or Section 6 of Annex X, as applicable, shall also apply.
11.
In addition to the procedures applicable pursuant to paragraph 3, 4, 6, or 7, in the case of devices that are composed of substances or of combinations of substances that are intended to be introduced into the human body via a body orifice or applied to the skin and that are absorbed by or locally dispersed in the human body, the procedure specified in Section 5.4 of Annex IX or Section 6 of Annex X, as applicable, shall also apply.
12.
The Member State in which the notified body is established may require that all or certain documents, including the technical documentation, audit, assessment and inspection reports, relating to the procedures referred to in paragraphs 1 to 7 and 9 to 11 be made available in an official Union language(s) determined by that Member State. In the absence of such requirement, those documents shall be available in any official Union language acceptable to the notified body.
13.
Investigational devices shall be subject to the requirements set out in Articles 62 to 81.
14.
The Commission may, by means of implementing acts, specify detailed arrangements and procedural aspects with a view to ensuring the harmonised application of the conformity assessment procedures by the notified bodies for any of the following aspects:
(a)
the frequency and the sampling basis of the assessment of the technical documentation on a representative basis as set out in the third paragraph of Section 2.3 and in Section 3.5 of Annex IX in the case of class IIa and class IIb devices, and in Section 10.2 of Annex XI in the case of class IIa devices;
(b)
the minimum frequency of unannounced on-site audits and sample tests to be conducted by notified bodies in accordance with Section 3.4 of Annex IX, taking into account the risk-class and the type of device;
(c)
the physical, laboratory or other tests to be carried out by notified bodies in the context of sample tests, assessment of the technical documentation and type examination in accordance with Sections 3.4 and 4.3 of Annex IX, Section 3 of Annex X and Section 15 of Annex XI.
The implementing acts referred to in the first subparagraph shall be adopted in accordance with the examination procedure referred to in Article 114(3).
Official catalogue · non-binding
Related guidance (22)
Current English catalogue entries that refer to Article 52. Guidance is non-binding and kept separate from the regulation text.
Unverified automated extraction preview. This text has not yet been checked against the official source and may contain omissions, ordering errors, table errors, missing images, or formatting errors. Consult the official source before relying on it.
Unofficial, non-binding reproduction. This reproduction is provided for convenience and does not create legal obligations. Consult the linked official source; the official source controls.
MDCG 2019-15 rev.1Guidance notes for manufacturers of class I medical devicesDecember 2019⌄
Unverified automated extraction preview. This text has not yet been checked against the official source and may contain omissions, ordering errors, table errors, missing images, or formatting errors. Consult the official source before relying on it.
Unofficial, non-binding reproduction. This reproduction is provided for convenience and does not create legal obligations. Consult the linked official source; the official source controls.
MDCG 2024-10Clinical evaluation of orphan medical devicesJune 2024⌄
Unverified automated extraction preview. This text has not yet been checked against the official source and may contain omissions, ordering errors, table errors, missing images, or formatting errors. Consult the official source before relying on it.
Unofficial, non-binding reproduction. This reproduction is provided for convenience and does not create legal obligations. Consult the linked official source; the official source controls.
MDCG 2020-6Guidance on sufficient clinical evidence for legacy devicesBackground note on the relationship between MDCG 2020-6 and MEDDEV 2.7/1 rev.4 on clinical evaluationApril 2020⌄
Unverified automated extraction preview. This text has not yet been checked against the official source and may contain omissions, ordering errors, table errors, missing images, or formatting errors. Consult the official source before relying on it.
Unverified automated extraction preview. This text has not yet been checked against the official source and may contain omissions, ordering errors, table errors, missing images, or formatting errors. Consult the official source before relying on it.
Unofficial, non-binding reproduction. This reproduction is provided for convenience and does not create legal obligations. Consult the linked official source; the official source controls.
MDCG 2019-9 - rev.1Summary of safety and clinical performanceMarch 2022⌄
Unverified automated extraction preview. This text has not yet been checked against the official source and may contain omissions, ordering errors, table errors, missing images, or formatting errors. Consult the official source before relying on it.
Unofficial, non-binding reproduction. This reproduction is provided for convenience and does not create legal obligations. Consult the linked official source; the official source controls.
MDCG 2021-3Questions and Answers on Custom-Made DevicesMarch 2021⌄
Unverified automated extraction preview. This text has not yet been checked against the official source and may contain omissions, ordering errors, table errors, missing images, or formatting errors. Consult the official source before relying on it.
Unofficial, non-binding reproduction. This reproduction is provided for convenience and does not create legal obligations. Consult the linked official source; the official source controls.
Gradual roll out of EUDAMEDQ&A on practical aspects related to the implementation of the gradual roll-out of Eudamed pursuant to the MDR and IVDR, as amended by Regulation (EU) 2024/1860 amending Regulations (EU) 2017/745 and (EU) 2017/746 as regards a gradual roll-out of Eudamed, the obligation to inform in case of interruption or discontinuation of supply, and transitional provisions for certain in vitro diagnostic medical devices1November 2024⌄
Gradual roll out of EUDAMEDPDF · Unverified extraction previewOfficial source
Unverified automated extraction preview. This text has not yet been checked against the official source and may contain omissions, ordering errors, table errors, missing images, or formatting errors. Consult the official source before relying on it.
Unofficial, non-binding reproduction. This reproduction is provided for convenience and does not create legal obligations. Consult the linked official source; the official source controls.
MDCG 2021-13 rev.1Questions and answers on obligations and related rules for the registration in EUDAMED of actors other than manufacturers, authorised representatives and importers subject to the obligations of Article 31 MDR and Article 28 IVDRJuly 2021⌄
Unverified automated extraction preview. This text has not yet been checked against the official source and may contain omissions, ordering errors, table errors, missing images, or formatting errors. Consult the official source before relying on it.
Unofficial, non-binding reproduction. This reproduction is provided for convenience and does not create legal obligations. Consult the linked official source; the official source controls.
Show all 22 guidance documentsShow only the first 8
MDCG 2021-1 rev.1Guidance on harmonised administrative practices and alternative technical solutions until EUDAMED is fully functionalMay 2021⌄
Unverified automated extraction preview. This text has not yet been checked against the official source and may contain omissions, ordering errors, table errors, missing images, or formatting errors. Consult the official source before relying on it.
Unofficial, non-binding reproduction. This reproduction is provided for convenience and does not create legal obligations. Consult the linked official source; the official source controls.
MDCG 2023-4Medical Device Software (MDSW) – Hardware combinations Guidance on MDSW intended to work in combination with hardware or hardware componentsOctober 2023⌄
Unverified automated extraction preview. This text has not yet been checked against the official source and may contain omissions, ordering errors, table errors, missing images, or formatting errors. Consult the official source before relying on it.
Unofficial, non-binding reproduction. This reproduction is provided for convenience and does not create legal obligations. Consult the linked official source; the official source controls.
MDCG 2019-16 rev.1Guidance on cybersecurity for medical devicesJuly 2020⌄
Unverified automated extraction preview. This text has not yet been checked against the official source and may contain omissions, ordering errors, table errors, missing images, or formatting errors. Consult the official source before relying on it.
Unofficial, non-binding reproduction. This reproduction is provided for convenience and does not create legal obligations. Consult the linked official source; the official source controls.
MDCG 2024-6Preliminary re-assessment review (PRAR) form template (MDR)May 2024⌄
Unverified automated extraction preview. This text has not yet been checked against the official source and may contain omissions, ordering errors, table errors, missing images, or formatting errors. Consult the official source before relying on it.
Unofficial, non-binding reproduction. This reproduction is provided for convenience and does not create legal obligations. Consult the linked official source; the official source controls.
MDCG 2024-7 rev.1 MDCG 2021-15/MDCG 2024-7 AnnexPreliminary assessment review (PAR) form template (MDR) Annex to Application Form & PAR Template MDR (List of documents)January 2025⌄
Unverified automated extraction preview. This text has not yet been checked against the official source and may contain omissions, ordering errors, table errors, missing images, or formatting errors. Consult the official source before relying on it.
Unverified automated extraction preview. This text has not yet been checked against the official source and may contain omissions, ordering errors, table errors, missing images, or formatting errors. Consult the official source before relying on it.
Unofficial, non-binding reproduction. This reproduction is provided for convenience and does not create legal obligations. Consult the linked official source; the official source controls.
MDCG 2022-4 rev.2Guidance on appropriate surveillance regarding the transitional provisions under Article 120 of the MDR with regard to devices covered by certificates according to the MDD or the AIMDDMay 2024⌄
Unverified automated extraction preview. This text has not yet been checked against the official source and may contain omissions, ordering errors, table errors, missing images, or formatting errors. Consult the official source before relying on it.
Unofficial, non-binding reproduction. This reproduction is provided for convenience and does not create legal obligations. Consult the linked official source; the official source controls.
MDCG 2019-6 rev.5Questions and answers: Requirements relating to notified bodiesFebruary 2025⌄
Unverified automated extraction preview. This text has not yet been checked against the official source and may contain omissions, ordering errors, table errors, missing images, or formatting errors. Consult the official source before relying on it.
Unofficial, non-binding reproduction. This reproduction is provided for convenience and does not create legal obligations. Consult the linked official source; the official source controls.
MDCG 2021-23Guidance for notified bodies, distributors and importers on certification activities in accordance with Article 16(4) of Regulation (EU) 2017/745 and Regulation (EU) 2017/746August 2021⌄
Unverified automated extraction preview. This text has not yet been checked against the official source and may contain omissions, ordering errors, table errors, missing images, or formatting errors. Consult the official source before relying on it.
Unofficial, non-binding reproduction. This reproduction is provided for convenience and does not create legal obligations. Consult the linked official source; the official source controls.
MDCG 2021-15 rev.1 MDCG 2021-15/MDCG 2024-7 AnnexApplication form to be submitted by a conformity assessment body when applying for designation as notified body under the medical devices regulation (MDR)Annex to Application Form & PAR Template MDR (List of documents)January 2025⌄
Unverified automated extraction preview. This text has not yet been checked against the official source and may contain omissions, ordering errors, table errors, missing images, or formatting errors. Consult the official source before relying on it.
Unverified automated extraction preview. This text has not yet been checked against the official source and may contain omissions, ordering errors, table errors, missing images, or formatting errors. Consult the official source before relying on it.
Unofficial, non-binding reproduction. This reproduction is provided for convenience and does not create legal obligations. Consult the linked official source; the official source controls.
MDCG 2020-14Guidance for notified bodies on the use of MDSAP audit reports in the context of surveillance audits carried out under the Medical Devices Regulation (MDR)/In Vitro Diagnostic medical devices Regulation (IVDR)August 2020⌄
Unverified automated extraction preview. This text has not yet been checked against the official source and may contain omissions, ordering errors, table errors, missing images, or formatting errors. Consult the official source before relying on it.
Unofficial, non-binding reproduction. This reproduction is provided for convenience and does not create legal obligations. Consult the linked official source; the official source controls.
MDCG 2020-12Guidance on transitional provisions for consultations of authorities on devices incorporating a substance which may be considered a medicinal product and which has action ancillary to that of the device, as well as on devices manufactured using TSE susceptible animal tissuesJune 2020⌄
Unverified automated extraction preview. This text has not yet been checked against the official source and may contain omissions, ordering errors, table errors, missing images, or formatting errors. Consult the official source before relying on it.
Unofficial, non-binding reproduction. This reproduction is provided for convenience and does not create legal obligations. Consult the linked official source; the official source controls.
MDCG 2019-13 rev.1Guidance on sampling of devices for the assessment of the technical documentationDecember 2024⌄
Unverified automated extraction preview. This text has not yet been checked against the official source and may contain omissions, ordering errors, table errors, missing images, or formatting errors. Consult the official source before relying on it.
Unofficial, non-binding reproduction. This reproduction is provided for convenience and does not create legal obligations. Consult the linked official source; the official source controls.
MDCG 2018-8Guidance on content of the certificates, voluntary certificate transfersNovember 2018⌄
Unverified automated extraction preview. This text has not yet been checked against the official source and may contain omissions, ordering errors, table errors, missing images, or formatting errors. Consult the official source before relying on it.
Unofficial, non-binding reproduction. This reproduction is provided for convenience and does not create legal obligations. Consult the linked official source; the official source controls.
MDCG 2022-18MDCG Position Paper on the application of Article 97 MDR to legacy devices for which the MDD or AIMDD certificate expires before the issuance of a MDR certificateDecember 2022⌄
Unverified automated extraction preview. This text has not yet been checked against the official source and may contain omissions, ordering errors, table errors, missing images, or formatting errors. Consult the official source before relying on it.
Unofficial, non-binding reproduction. This reproduction is provided for convenience and does not create legal obligations. Consult the linked official source; the official source controls.
Official EU legal act
Decision No 768/2008/EC of the European Parliament and of the Council of 9 July 2008 on a common framework for the marketing of products
Select the citation to open the authentic EUR-Lex version.
Official EU legal act
Regulation (EC) No 1272/2008 of the European Parliament and of the Council of 16 December 2008 on classification, labelling and packaging of substances and mixtures, amending and repealing Directives 67/548/EEC and 1999/45/EC, and amending Regulation (EC) No 1907/2006
Select the citation to open the authentic EUR-Lex version.
Official EU legal act
Council Directive 80/181/EEC of 20 December 1979 on the approximation of the laws of the Member States relating to units of measurement and on the repeal of Directive 71/354/EEC
Select the citation to open the authentic EUR-Lex version.
Official EU legal act
Regulation (EC) No 1907/2006 of the European Parliament and of the Council of 18 December 2006 concerning the Registration, Evaluation, Authorisation and Restriction of Chemicals (REACH), establishing a European Chemicals Agency, amending Directive 1999/45/EC and repealing Council Regulation (EEC) No 793/93 and Commission Regulation (EC) No 1488/94 as well as Council Directive 76/769/EEC and Commission Directives 91/155/EEC, 93/67/EEC, 93/105/EC and 2000/21/EC
Select the citation to open the authentic EUR-Lex version.
Official EU legal act
Council Directive 80/181/EEC of 20 December 1979 on the approximation of the laws of the Member States relating to units of measurement and on the repeal of Directive 71/354/EEC
Select the citation to open the authentic EUR-Lex version.
Official EU legal act
Council Directive 85/374/EEC of 25 July 1985 on the approximation of the laws, regulations and administrative provisions of the Member States concerning liability for defective products
Select the citation to open the authentic EUR-Lex version.
Official EU legal act
Regulation (EU) 2017/745 of the European Parliament and of the Council of 5 April 2017 on medical devices, amending Directive 2001/83/EC, Regulation (EC) No 178/2002 and Regulation (EC) No 1223/2009 and repealing Council Directives 90/385/EEC and 93/42/EEC
Select the citation to open the authentic EUR-Lex version.
Official EU legal act
Regulation (EC) No 1907/2006 of the European Parliament and of the Council of 18 December 2006 concerning the Registration, Evaluation, Authorisation and Restriction of Chemicals (REACH), establishing a European Chemicals Agency, amending Directive 1999/45/EC and repealing Council Regulation (EEC) No 793/93 and Commission Regulation (EC) No 1488/94 as well as Council Directive 76/769/EEC and Commission Directives 91/155/EEC, 93/67/EEC, 93/105/EC and 2000/21/EC
Select the citation to open the authentic EUR-Lex version.
Official EU legal act
Council Directive 93/42/EEC of 14 June 1993 concerning medical devices
Select the citation to open the authentic EUR-Lex version.
Official EU legal act
Directive 2006/42/EC of the European Parliament and of the Council of 17 May 2006 on machinery, and amending Directive 95/16/EC
Select the citation to open the authentic EUR-Lex version.
Official EU legal act
Directive 98/79/EC of the European Parliament and of the Council of 27 October 1998 on in vitro diagnostic medical devices
Select the citation to open the authentic EUR-Lex version.
Official EU legal act
Regulation (EC) No 1907/2006 of the European Parliament and of the Council of 18 December 2006 concerning the Registration, Evaluation, Authorisation and Restriction of Chemicals (REACH), establishing a European Chemicals Agency, amending Directive 1999/45/EC and repealing Council Regulation (EEC) No 793/93 and Commission Regulation (EC) No 1488/94 as well as Council Directive 76/769/EEC and Commission Directives 91/155/EEC, 93/67/EEC, 93/105/EC and 2000/21/EC
Select the citation to open the authentic EUR-Lex version.
Official EU legal act
Regulation (EU) No 536/2014 of the European Parliament and of the Council of 16 April 2014 on clinical trials on medicinal products for human use, and repealing Directive 2001/20/EC
Select the citation to open the authentic EUR-Lex version.
Official EU legal act
Regulation (EU) 2017/745 of the European Parliament and of the Council of 5 April 2017 on medical devices, amending Directive 2001/83/EC, Regulation (EC) No 178/2002 and Regulation (EC) No 1223/2009 and repealing Council Directives 90/385/EEC and 93/42/EEC
Select the citation to open the authentic EUR-Lex version.
Official EU legal act
Directive 2002/98/EC of the European Parliament and of the Council of 27 January 2003 setting standards of quality and safety for the collection, testing, processing, storage and distribution of human blood and blood components
Select the citation to open the authentic EUR-Lex version.
Official EU legal act
Directive 2004/10/EC of the European Parliament and of the Council of 11 February 2004 on the harmonisation of laws, regulations and administrative provisions relating to the application of the principles of good laboratory practice and the verification of their applications for tests on chemical substances
Select the citation to open the authentic EUR-Lex version.
Official EU legal act
Directive 2004/23/EC of the European Parliament and of the Council of 31 March 2004 on setting standards of quality and safety for the donation, procurement, testing, processing, preservation, storage and distribution of human tissues and cells
Select the citation to open the authentic EUR-Lex version.
Official EU legal act
Directive 2006/42/EC of the European Parliament and of the Council of 17 May 2006 on machinery, and amending Directive 95/16/EC
Select the citation to open the authentic EUR-Lex version.
Official EU legal act
Directive 2010/63/EU of the European Parliament and of the Council of 22 September 2010 on the protection of animals used for scientific purposes
Select the citation to open the authentic EUR-Lex version.
Official EU legal act
Directive 2011/83/EU of the European Parliament and of the Council of 25 October 2011 on consumer rights
Select the citation to open the authentic EUR-Lex version.
Official EU legal act
Directive 2014/30/EU of the European Parliament and of the Council of 26 February 2014 on the harmonisation of the laws of the Member States relating to electromagnetic compatibility
Select the citation to open the authentic EUR-Lex version.
Official EU legal act
Directive 2014/35/EU of the European Parliament and of the Council of 26 February 2014 on the harmonisation of the laws of the Member States relating to the making available on the market of electrical equipment designed for use within certain voltage limits
Select the citation to open the authentic EUR-Lex version.
Official EU legal act
Directive 2014/53/EU of the European Parliament and of the Council of 16 April 2014 on the harmonisation of the laws of the Member States relating to the making available on the market of radio equipment
Select the citation to open the authentic EUR-Lex version.
Official EU legal act
Regulation (EU) 2017/745 of the European Parliament and of the Council of 5 April 2017 on medical devices, amending Directive 2001/83/EC, Regulation (EC) No 178/2002 and Regulation (EC) No 1223/2009 and repealing Council Directives 90/385/EEC and 93/42/EEC
Select the citation to open the authentic EUR-Lex version.
Official EU legal act
Regulation (EC) No 726/2004 of the European Parliament and of the Council of 31 March 2004 laying down Community procedures for the authorisation and supervision of medicinal products for human and veterinary use and establishing a European Medicines Agency
Select the citation to open the authentic EUR-Lex version.
Official EU legal act
Regulation (EC) No 1272/2008 of the European Parliament and of the Council of 16 December 2008 on classification, labelling and packaging of substances and mixtures, amending and repealing Directives 67/548/EEC and 1999/45/EC, and amending Regulation (EC) No 1907/2006
Select the citation to open the authentic EUR-Lex version.
Official EU legal act
Regulation (EC) No 1272/2008 of the European Parliament and of the Council of 16 December 2008 on classification, labelling and packaging of substances and mixtures, amending and repealing Directives 67/548/EEC and 1999/45/EC, and amending Regulation (EC) No 1907/2006
Select the citation to open the authentic EUR-Lex version.
Official EU legal act
Regulation (EC) No 765/2008 of the European Parliament and of the Council of 9 July 2008 setting out the requirements for accreditation and market surveillance relating to the marketing of products
Select the citation to open the authentic EUR-Lex version.
Official EU legal act
Regulation (EU) 2017/745 of the European Parliament and of the Council of 5 April 2017 on medical devices, amending Directive 2001/83/EC, Regulation (EC) No 178/2002 and Regulation (EC) No 1223/2009 and repealing Council Directives 90/385/EEC and 93/42/EEC
Select the citation to open the authentic EUR-Lex version.
Official EU legal act
Regulation (EU) No 1025/2012 of the European Parliament and of the Council of 25 October 2012 on European standardisation
Select the citation to open the authentic EUR-Lex version.
Official EU legal act
Regulation (EU) No 528/2012 of the European Parliament and the Council of 22 May 2012 concerning the making available on the market of and use of biocidal products
Select the citation to open the authentic EUR-Lex version.
Official EU legal act
Regulation (EU) No 536/2014 of the European Parliament and of the Council of 16 April 2014 on clinical trials on medicinal products for human use, and repealing Directive 2001/20/EC
Select the citation to open the authentic EUR-Lex version.
Official EU legal act
Regulation (EU) 2016/679 of the European Parliament and of the Council of 27 April 2016 on the protection of natural persons with regard to the processing of personal data and on the free movement of such data (General Data Protection Regulation)
Select the citation to open the authentic EUR-Lex version.
Official EU legal act
Regulation (EU) 2017/745 of the European Parliament and of the Council of 5 April 2017 on medical devices, amending Directive 2001/83/EC, Regulation (EC) No 178/2002 and Regulation (EC) No 1223/2009 and repealing Council Directives 90/385/EEC and 93/42/EEC
Select the citation to open the authentic EUR-Lex version.
Official EU legal act
Regulation (EU) 2017/746 of the European Parliament and of the Council of 5 April 2017 on in vitro diagnostic medical devices
Select the citation to open the authentic EUR-Lex version.
Official EU legal act
Regulation (EU) 2022/123 of the European Parliament and of the Council of 25 January 2022 on a reinforced role for the European Medicines Agency in crisis preparedness and management for medicinal products and medical devices
Select the citation to open the authentic EUR-Lex version.